Setting MotionAlert Access
Setting this parameter allows the user to access the MOTIONALERT page from the MEASURE page.
If access is denied here:
-
the user cannot access the MOTIONALERT page to enable or disable that feature
-
the InstantAlert feature (chapter 2.3) cannot be activated.
(1)
To grant or deny user access to the MOTIONALERT page, use the button to change the indi-
cated selection.
User access is:
permitted when the setting indicates ON.
denied when the setting indicates OFF.
(2)
The selection is confirmed by pressing either the  or  button.
Setting Sensor Alarms
This page allows modifying the preset alarm values of:
-
LOW Alarm
-
HIGH Alarm
-
STEL Alarm
-
TWA Alarm.
Factory set alarm levels are shown in chapter 5.1.
(1)
Press the q button to enter Sensor Alarm setup.
LOW Alarm Setup screen displays.
Fig. 8
Sensor Alarm Setup
(2)
Press
the  button to abort the operation or
the  button to go to next alarm setup or
the q button to change the alarm setpoints.
Alarm Value for the first Sensor displays.
ALTAIR 5X/ALTAIR 5X IR
Operation
US
34