IEC 60601-1-2:2007 Table 4 – Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The XV1 is intended for use in the electromagnetic environment specified below. The customer or user of the XV1
should assure that it is used in such an environment
Immunity test
IEC 60601 test level
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
a
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the XV1 is used exceeds
the applicable RF compliance level above, the XV1 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the XV1.
Orascoptic XV1—Instructions for Use
Compliance level
3 V/m
80 MHz to 2.5 GHz
U.S. Technical Support: 800-369-3698
Outside the U.S.—Call Your Local Distributor
Electromagnetic environment – guidance
d = 1.167 √P 80 MHz to 800 MHz
d = 2.33 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m). Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
should be less than the compliance
a
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol: