Selecting the location Page 7
Switching on the balance Page 14
Leveling the balance Page 14
Performing an internal adjustment Page 14
4.7.2 Transporting the balance over long distances
METTLER TOLEDO recommends using the original packaging for transportation or shipment of the balance or
balance components over long distances. The elements of the original packaging are developed specifically for
the balance and its components and ensure maximum protection during transportation.
See also
Unpacking the balance Page 8
4.7.3 Packing and storing
Packing the balance
Store all parts of packaging in a safe place. The elements of the original packaging are developed specifically
for the balance and its components, and ensures maximum protection during transportation and storage.
Storing the balance
Only store the balance under the following conditions:
Indoor and in the original packaging
According to the environmental conditions, see "Technical Data"
Note
When storing for longer than 6 months, the rechargeable battery may become empty (only date and time get
lost).
See also
Technical Data Page 21
5 Maintenance
To guarantee the functionality of the balance and the accuracy of the weighing results, a number of
maintenance actions must be performed by the user.
For further information, consult the Reference Manual (RM).
5.1 Maintenance tasks
Maintenance action
Performing an internal
adjustment
Performing routine tests
(eccentricity test, repeata-
bility test, sensitivity test).
METTLER TOLEDO
recommends to at least
perform a sensitivity test.
Pipette Calibration Balances
Recommended interval
Daily
After cleaning
After leveling
After changing the location
After cleaning
After assembling the balance
After a software update
Depending on your internal regulations
(SOP)
u
www.mt.com/XPR26PC-RM
Remarks
see "Performing an internal
adjustment"
see "Tests" in the Reference
Manual
17