5.3 Cleaning
5.3.1 Disassembling for cleaning
The containers are empty, see [Emptying the containers Page 18].
The balance is switched off, see [Switching off the balance Page 15].
1 Remove the fixing screw (1) and put the light barrier (2)
aside.
2 Remove the evaporation trap door (3).
3 Carefully remove the evaporation trap cover (4).
4 Remove the centering ring (5).
5 Carefully lift and remove the pipetting base (6) with the
evaporation trap container (7).
6 Remove the pipetting container (8) and unscrew the
cover holding the pipetting tube (9).
7 Carefully remove the pipetting container support (10).
8 Remove the drip tray (11).
5.3.2 Cleaning the balance
NOTICE
Damage to the instrument due to inappropriate cleaning methods
If liquid enters the housing, it can damage the instrument. The surface of the instrument can be
damaged by certain cleaning agents, solvents, or abrasives.
1 Do not spray or pour liquid on the instrument.
2 Only use the cleaning agents specified in the Reference Manual (RM) of the instrument or the
3 Only use a lightly moistened, lint-free cloth or a tissue to clean the instrument.
4 Wipe off any spills immediately.
For further information on cleaning a balance, consult "8 Steps to a Clean Balance".