Warnings
To avoid damage to the DI water pump, ensure that the DI water
reservoir is filled before turning on the Analyzer.
To avoid false TOC readings and possible damage to the Analyzer,
always make sure the sample inlet is open and the DI water reservoir is
filled before turning on the Analyzer.
Always close the door of the iOS system when not sampling from a vial.
To protect the Analyzer from damage, always drain the water from the
Analyzer prior to shipping.
When cleaning the Analyzer, always apply liquid to the cloth directly,
and do not spray liquids directly on the Analyzer.
Do not use water or cleaners on the touchscreen, as water can damage
or discolor the polarizer. Refer to the next section for recommendations
on cleaning the touchscreen.
Wipe away any condensation immediately.
Operating the Analyzer in Online mode with no sample flow and with
the Flow Sensor off can result in erroneous measurements.
Standards for the Sievers M500 TOC Analyzer are acidified. If
preparing your own standards and using glass vials, ensure to acidify
the standards to a pH of 3, or the system protocols will yield poor
results.
Ensure to center the vent tube within the tubing fittings to prevent
pinching beneath the vent tube and prevent the sample from flowing.