SECTION 8—TROUBLESHOOTING GUIDE
SECTION 8—TROUBLESHOOTING
GUIDE
If your XPO
Portable Oxygen Concentrator fails to operate
2
properly, refer to the following chart for possible causes and
solutions. If necessary, contact your service provider.
ALARM/WARNING:
SYMPTOM
START-UP INDICATOR
XPO
does not
2
operate when
On/Off button
pressed.
OR
Front panel lights
DO NOT
illuminate upon
start-up
OR
All front panel
lights go dark.
NO BREATH DETECTED ALARM:
Continuous
audible beep and
red alarm
indicator is on
continuously (not
blinking) upon
start up.
Invacare XPO2™
PROBABLE CAUSE SOLUTION
On/Off button was not held
down long enough.
Internal battery is
discharged (and
supplemental battery if
connected to the unit).
Malfunction
XPO
has not detected a
2
breath for a predetermined
time period.
40
Try to power up the
unit again while
continuing to press the
On/Off button until the
front panel lights begin
to illuminate. This takes
typically 2-3 seconds.
Connect XPO
to either
2
the AC or DC charger
and retry.
Change to an alternate
oxygen supply and
contact your equipment
provider.
1. Verify the cannula is
connected, not kinked,
properly positioned and
you are breathing
through your nose.
2. If the alarm continues,
change to another
source of oxygen and
contact your equipment
provider.
Part No 1160887