rainbow® Series Patient Cables, EMS
For Pulse Oximeters with rainbow Technology
Reusable
INDICATIONS
The rainbow® Series Patient Cables, EMS have the same indications for use as the applicable sensors. Please see the compatible
sensor's directions for use for indications for use and prescribed information.
DESCRIPTION
The rainbow Series Patient Cables, EMS are used with Masimo® rainbow and M-LNCS® Sensors.rainbow Series Patient Cables,
EMS are for use only with devices containing Masimo rainbow SET™ technology Version 7.1 or higher or licensed to use
rainbow compatible sensors. Consult individual oximetry system manufacturers for compatibility of particular sensor and
device models. Each device manufacturer is responsible for determining whether their devices are compatible with each
sensor model.
rainbow Series Patient Cables, EMS were verified using Masimo rainbow SET technology.
WARNING: Masimo sensors and cables are designed for use with devices containing Masimo SET® or Masimo rainbow SET
oximetry or licensed to use Masimo sensors.
WARNINGS, CAUTIONS, AND NOTES
• Always refer to the oximeter module operator's manual for complete instructions or additional instructions.
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable and
discard if cracks or discoloration are found.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable and
sensor before use, otherwise degraded performance and/or patient injury can result.
• Carefully route patient cable to reduce the possibility of patient entanglement or strangulation.
• Failure to properly connect the sensor or the oximeter module to the cable will result in intermittent readings, inaccurate
results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or disconnecting
either end.
• To
prevent
damage,
Do not attempt to sterilize the cable or sensor.
• Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may damage
the electrical components, potentially leading to patient harm.
• Caution: Replace the cable when a replace cable message is displayed, or when a low SIQ message is consistently
displayed while monitoring consecutive patients after completing the low SIQ troubleshooting steps identified in the
monitoring device operator's manual.
• Note: The cable is provided with X-Cal® technology to minimize the risk of inaccurate readings and unanticipated loss
of patient monitoring. The cable will provide up to 17,520 hours of patient monitoring time. Replace the cable when the
patient monitoring time is exhausted. Some legacy monitoring devices do not distinguish between the cable and the
sensor. In these devices, a replace sensor or similar message could pertain either to the sensor or the cable.
CABLE AND SENSOR SELECTION
NOTE: The maximum length for the sensor, cable and extension cable combination is 15 feet.
Refer to the table below for correct examples:
Cable
Length
1 ft
4 ft
12 ft
INSTRUCTIONS
A. Connect the rainbow patient cable, EMS to the Device
1. Connect the device connector of the patient cable to the mating connector of the device.
2. Ensure the patient cable connector is fully engaged with the device. A tactile or audible click of connection should be
heard.
B. Connect the rainbow patient cable, EMS to the Sensors
Connect the rainbow patient cable, EMS to an M-LNCS disposable sensor:
1. Refer to Fig. 1. Orient the sensor connector to the patient cable connector as shown.
2. Refer to Fig. 2. Insert the sensor connector completely into the patient cable connector until it locks in place as shown.
D IREC TIO N S FO R US E
Not made with natural rubber latex
LATEX
PCX-2108A
02/13
do
not
soak
or
18 inches
Yes
Yes
No
immerse
the
sensor
Sensor
3 feet
Yes
Yes
No
3
or
cable
in
any
6 feet
Yes
No
No
en
Non-Sterile
liquid
solution.