SpO
Lo-Limit: SpO
2
the step is 1%. The factory default value for adult is 90%
and 95% for pediatric.
PR Hi-Limit: High limit setting of pulse rate; range:
100~240bpm. From 100 to 150, the step is 1bpm, and
from 150 to 240, the step is 5bpm. The factory default
value for adult is 120bpm and 160bpm for pediatric.
PR Lo-Limit: Low limit setting of pulse rate; range:
30~99bpm, and the step is 1bpm. The factory default
value for adult is 50bpm and 60bpm for pediatric.
Note: When the SpO2 reading is equal to or below the preset
alert setting or the PR reading is equal to or above the preset
alert setting, then the over-limit alert sound will trigger, and the
corresponding reading(s) will blink. When measured on
pediatric patients, the alert sound and blinking display will
trigger if the SpO2 reading is off the limit for 10 seconds.