Operating Humidity:
Atmospheric pressure:
Note: portable and mobile RF communications equipment
may affect the performance of the Oximeter.
G. Data Update
8 beats moving average for both SpO
H. Data Storage
Recording SpO
up to 384–hour records can be stored.
I. Low Perfusion Performance
The accuracy of SpO
precision described above when the modulation amplitude
is at least 0.5%.
J. Resistance to interference of surrounding light:
The difference between the SpO
condition of indoor natural light and that of darkroom is less
than ±1%.
K. Dimensions: 145 mm (L) × 74 mm (W) × 29 mm (H)
Net Weight: 210g (including batteries)
L. Classification
Type of protection against electric shock:
Internally powered equipment
Degree of protection:
Type BF applied parts.
Degree of protection against harmful ingress of liquids:
IP22 .
User Manual for Pulse Oximeter
30%~80%
70kPa~106kPa
and pulse rate data every 1/2/4/8 second(s),
2
and PR measurement will meet the
2
22
and pulse rate readings
2
value measured in the
2