5. Setting the treatment values
Warning!
m
Patient hazard –
Before therapy, a test run consisting
of several exercise cycles must be com‑
pleted without the patient. Then repeat
the test run with the patient and check
that the movement does not cause any
pain.
Note: See also 2.2 and 2.3 as well as
page V!
Note!
Programming is only possible when the
patient's chip card has been inserted.
For information about therapy param‑
eters and programming of special func‑
tions, please refer to sections 5.1 to 5.5.
See section 5.6 for programming
examples.
Important!
The programming unit ARTROMOT®-S3
"Graphics" can be connected to all
products of the ARTROMOT® shoulder
series with serial numbers 3000 and
higher.
It is possible to exchange chipcards
between the "Text" and "Graphics"
versions.
Please note:
1. If you use a "Graphics" programming
unit with a formatted chipcard of
the "Text" version, please note the
following
– the selected language of the"Text"
version remains activated"; it is ir‑
relevant for the "Graphics" version
– the reverse on load setting saved
in the "Text" version is automati‑
cally used for both motors
– the key lock is disabled.
2. If you use a "Text" programming unit
with a formatted chipcard of the
"Graphics" version, please note the
following
– the selected language automati‑
cally changes to the default set‑
ting: German
– the reverse on load setting saved
for motor A in the "Graphics" ver‑
sion is automatically used for both
motors
– a key lock set in the "Graphics" ver‑
sion will remain active, because it
is irrelevant for the "Text" version.
5.1 General information
on programming
ARTROMOT®-S3
1. You activate the programming mode by
briefly pressing the MENU key on the
programming unit.
2. The treatment parameters and func‑
tions are allocated to four (Standard
58