6.3 Transport
The following operating steps must
be completed before transporting the
ARTROMOT®‑S3:
1.
Activate the „transport setting"
function in the menu (see also 5.3) and
start the ARTROMOT®‑S3.
2.
Turn off the ARTROMOT®‑S3 power
switch (7).
3.
Remove the power cord (6) and dis‑
connect the motion element and the
programming unit (9).
4. Open locking screw (14) and pull out
the armrest (4) for the healthy arm.
5.
Grasping the holding tube (17) of the
motion element, open locking screw
(14) and pull out the motion element.
6. Set horizontal adduction/abduction to
0° (indexing knob 12).
7.
Open the locking screw for the head‑
rest adjustment (30) and remove the
headrest¹.
8. Open the locking screw for the shoul‑
der restraint (31) and remove¹.
9. Open the locking screw for the height
adjustment of the patient restraint (32)
and remove
.
1
10. Open the locking screw for adjustment
of the backrest (10), fold the backrest
to the front all the way until flat on the
seat and tighten the screw.
11. Remove the two locking pins (34). Pull
out the legs towards the sides, reverse
and reinsert. Then put the locking pins
back in place.
12. Only use the original shipping box for
transporting the device. DJO can not be
held liable for transport damage, if the
device was not shipped in its original
shipping box.
13. Place the ARTROMOT®‑S3 – legs down
– on the bottom of the box. The mark‑
ings on the bottom of the box indicate
the correct position.
14. Then put the enclosed polystyrene
block on the chair, observing the cut‑
outs in the block.
15. Put the programming unit (1) in the
supplied box. Place the motion ele‑
ment, the armrest, the components
of the neck and shoulder restraint and
the power cord in the respective cut‑
outs in the polystyrene block.
16. Then close the box.
Reverse the above steps to reassemble the
device after transport.
Warning!
m
Shock hazard –
Before use allow the ARTROMOT®‑S3
to reach room temperature. If the device
has been transported at temperatures
below 0 °C (32 °F), leave it to dry at
room temperature for about 2 hours,
until any condensation has disappeared.
¹Only with ARTROMOT®-S3 comfort
73