4. Rotate the gauge to view each measurement. First
locate the total femoral tunnel value. Rotate the
gauge to locate the desired graft in tunnel value.
The corresponding implant loop size is located to
the right of the graft in tunnel value in the gauge
body window. Rotate the gauge again to read the
corresponding femoral socket drill depth value.
5. Drill the femoral socket of appropriate diameter to
the depth indicated on the gauge.
6. Using the 4.5mm cannulated reamer, create a
tunnel over the passing pin such that it breaks
through the lateral femoral cortex.
7. Insert the implant as described in IFU 110962,
RIGIDLOOP Cortical Fixation Implant.

CLEANING AND STERILIZATION

These cleaning and sterilization instructions have
been validated for preparing reusable DePuy Mitek
instruments for reuse. It is the responsibility of the
end user to ensure that the cleaning and sterilization
is actually performed using appropriate equipment,
materials, and personnel to achieve the desired
result. This normally requires validation and routine
monitoring of the process. Any deviation from these
instructions should be evaluated for effectiveness and
potential adverse consequences.
The Variable Depth Gauge is packaged NON-
STERILE and comes assembled for use. After each
use, disassemble the device and clean according to
the following instructions.
Disassembling the Depth Gauge
1. Remove the probe from the gauge body. Slide the
probe away from the body until you feel resistance.
Apply an additional pull until you are able to
disengage the probe from the body (Figure 1).