Disinfect creased tube and adapter
Fit double tube
Branched end
of double tube
Grey adapter
Double tube
Sleeve
Transparent adapter
After disinfecting
Dry the assembled tube system completely with the aid of the therapy device. The
procedure can be found in the instructions for use for your therapy device.
6.3 Change of patient
If there is a change of patient, replace the tube system and clean the outside of the
VENTI-O
by wiping down with disinfectant.
2
2. Clean the double tube with a little disinfectant in
hot water so that there are no residues.
3. After disinfecting, rinse the double tube thoroughly
with distilled water.
4. Leave the parts to dry completely.
1. Proceed as described in "Clean the tube system".
2. After disinfecting, rinse thoroughly with distilled
water.
1. Push the short part of the double tube through the
transparent adapter and then push it as far as
possible onto the gray adapter.
2. Push the sleeve back onto the gray adapter.
3. Push the long part of the double tube into the
groove of the transparent adapter. The two ends
of the tube must lie smoothly next to one another.
4. Insert the free end of the double tube in the
creased tube.
5. Push the sleeve of the creased tube onto the trans-
parent adapter.
Sleeve
Hygiene preparation
15