11. Technical Data
11.1 Specifications
Product class under 93/42/EEC
Dimensions WxHxD in cm
Weight
Temperature range
– Operation
– Storage
Air pressure range
Pneumatic connection
Inlet:
Outlet:
Permitted humidity for operation and storage
Electrical connection
Current consumption during operation
Classification according to EN 60601-1
– Protection against electric shock
– Degree of protection against electric shock
Electromagnetic compatibility according to
EN 60601-1-2
– Radio interference suppresion
– Radio interference resistance
Max. permitted flow
Min. required O
supply pressure at valve
2
inlet for 4 l/min O
2
Max. permitted static pressure of pressure
source
Permitted
therapy pressures
We reserve the right to make design changes.
20
EN
Technical Data
VENTI-O
IIa
8.0 x 9.3 x 3.0
approx. 195 g
+5°C to +35°C
–40°C to +70°C
600 – 1100 hPa
only tubes with 4 mm internal diam.
only tube system WM 23737
≤ 95% r.h. (no dew formation)
12 V DC
125 mA
Class II
Type B
EN 55011
EN 61000-4-2 to 3,
EN 61000-4-8
≤ 4 l/min O
250 hPa
7000 hPa
4 to 35 hPa
(1hPa = 1mbar ≈1 cm H
2
2
O)
2