Instructions for use
Make sure that the reading position you want to
use is configured with the incubation program
corresponding to the biological indicator you
want to analyze.
Do not remove or change placement of BI once
the reading process has begun. If this occurs,
results may be invalidated.
Use a non-sterilized biological indicator as a
positive control every time a processed indica-
tor is incubated. For more information refer to
the Positive Control section of this user guide.
6-Place the rapid BI in a reading position and wait for
the result. Reading will start automatically. Immedia-
tely after placing the BI in a position, the auto-reader
IC10/20FRLCD will emit an audible alarm to indicate
that the readout started and will show the informa-
tion screen of the initiated position. Status position
light will turn violet and will blink until the result is
informed.
NOTE: If the auto-reader does not run automatically,
it might indicate that the auto-reader temperature is
not steady yet (check the temperature indicating li-
ght to remain steady) or that the incubation position
is disabled due to an error in the initial autotest.
When a BI is removed from it's position du-
ring incubation process, an audible alarm will
sound to indicate the tube should be returned
to the corresponding position.
If the indicator isn't re-introduced into the rea-
ding position after 10 seconds of being with-
drawn, the reading will be cancelled and the
cancelation will be informed through the prin-
ting of a ticket.
7-When a positive result is detected, the red light and
the audible alarm will turn on, this will indicate that
the sterilization process to which the BI was subject,
has failed. 30 seconds after removing the IB from the
incubation position, the alarm and the red light will
turn off automatically and the position will be availa-
ble to start a new reading.
The negative BIs will be informed with a green light in
the corresponding position, this will indicate that the
sterilization process to which it was subject has been
successful. The green light will remain on for up to 30
seconds after the BI is withdrawn from the reading po-
sition. The auto-reader will inform results within 20 min.
(60 °C), 30 min. (60 ºC), 1 h. (60 ºC), 2 hs. (60 °C), 3 hs. (60
ºC) or 4 hs. (37 ºC) , according to the program selected.
These periods of incubation are the time-limits used by
the equipment to inform the fluorescence final readout.
Every time a positive result is detected, it will be infor-
med through the printing of the corresponding ticket
and with an audible alarm. The alarm can be cance-
lled by pressing the button
8-To register the tickets, press the button for paper
traction, take the paper and pull it up to cut it with
the serrated edge of the printer.
NOTE: If the paper runs out, the light of paper repla-
.
IC10/20FRLCD Page 8
cement will start blinking (see Replacement of the pa-
per roll in the printer).

Thermal paper specifications

Recommended paper: JUJO AF50KSE3 or similar (or-
der code ICTP).
Paper width : 58 mm
Maximum paper thickness: 60 g/m2
Maximum diameter size: 50 mm

Replacement of the paper roll

The printer uses thermal paper rolls.
1. Pull the handle of the
printer's door.
2. Open the printer's
cover and remove the
spent roll.
3. Place the new paper
roll with the outer side
up.
4. Close the printer co-
ver by pressing on the
sides of the lid.