ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration - electromagnetic emissions
The [KD-735] is intended for use in the electromagnetic environment specified below.
The customer or the user of the [KD-735] should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Not applicable
IEC 61000-3-3
226
Visiomed - KD-735 - 032016
Electromagnetic environment -
guidance
The [KD-735] uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The [KD-735] is suitable for use in all
establishments other than domestic and
those directly connected to the public low-
voltage power supply network that sup-
plies buildings used for domestic purposes.
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration - electromagnetic immunity
The [KD-735] is intended for use in the electromagnetic environment specified below. The
customer or the user of the [KD-735] should assure that it is used in such an environment.
IMMUNITY
IEC 60601test
test
level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Power
frequency
(50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
Visiomed - KD-735 - 032016
Compliance
Electromagnetic
level
environment - guidance
Floors should be wood, con-
crete or ceramic tile. If floors
± 6 kV contact
are covered with synthetic
± 8 kV air
material, the relative humid-
ity should be at least 30 %.
Power frequency magnetic
fields should be at levels char-
3 A/m
acteristic of a typical location
in a typical commercial or
hospital environment.
227