Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 36

TROUBLESHOOTING

If you have one of the following problems, please refer to
this breakdown service guide to help resolve the problem.
If the problem persists, please contact our customer service
at +33 892 350 334.
PROBLEM
POSSIBLE CAUSE
The cuff position
was not correct or
it was not properly
tightened
Body posture was
not correct during
testing
LCD Display shows
Speaking, arm or
abnormal result
body movement,
angry, excited or
nervous during
testing
Irregular heartbeat
(arrhythmia)
44
Visiomed - KD-735 - 032016
SOLUTION
Apply the cuff correctly
and try again
Review the "BODY
POSTURE DURING
MEASUREMENT" sections
of the instructions and
re-test.
Re-test when calm and
without speaking or
moving during the test
It is inappropriate
for people with
serious arrhythmia
to use this Electronic
Sphygmomanometer.
LCD shows low
Low Battery
battery symbol
Pressure system
LCD shows "Er 0"
is unstable before
measurement
Fail to detect
LCD shows "Er 1"
systolic pressure
Fail to detect
LCD shows "Er 2"
diastolic pressure
Pneumatic system
blocked or cuff is
LCD shows "Er 3"
too tight during
inflation
Pneumatic system
leakage or cuff is
LCD shows "Er 4"
too loose during
inflation
Visiomed - KD-735 - 032016
Change the batteries
Don't move and try again.
Apply the cuff correctly
and try again
45

Publicidad

Tabla de contenido
loading

Tabla de contenido