VISIOMED TensioFlash KD-735 Manual De Utilización página 116

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 36
230
Visiomed - KD-735 - 032016
Table 4
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-
SUPPORTING
Recommended separation distances between
Portable and mobile RF communications equipment and the [KD-735]
The [KD-735] is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the [KD-735] can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the [KD-735] as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
150 kHz to 80 MHz
d = 1.2 √P
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the recommended separa-
tion distance d in meters (m) can be determined using the equation applicable to the frequency
of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Visiomed - KD-735 - 032016
Visiomed - KD-735 - 072014
Separation distance according to
frequency of transmitter
m
80 MHz to 800 MHz
800 MHz to 2,5 GHz
d = 1.2 √P
d = 2.3 √P
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23
231

Publicidad

Tabla de contenido
loading

Tabla de contenido