Cleaning The Head; Arm Cleaning; Sterilising The Handles; Removing The Handles - DKL CHAIRS L2-S600 Manual Del Usuario

Ocultar thumbs Ver también para L2-S600:
Tabla de contenido

Publicidad

6.2
Cleaning must be carried out using a soft cloth in cotton or absorbent cotton with ethyl alcohol
or the specific PERFLEX detergent. Water-alcohol based disinfectants are suitable with 70%
isopropyl alcohol or ethanol.
6.3
Always use a cloth soaked in disinfectant approved to disinfect the surfaces and pass it over.

7. STERILISING THE HANDLES

7.1
To remove the handle, unscrew knob "A" and remove it from the support.
7.2
Before sterilising the handles, they must be decontaminated and disinfected.
To disinfect, Faro has tested the following products for disinfection: Eco Jet-1 (Cattani Group)
/ Sporekin Plus DS (Ims srl) / Zefirol Quick (Molteni Dental) / Durr FD366 Sensitive
62

CLEANING THE HEAD

Warning against danger of wear of the plastic and falling suspended weights
N ever spray detergent directly on the head. Do not use detergents-disinfectants
containing the following to clean plastic parts:
• AMMONIUM HYDROXIDE
• SODIUM HYDROXIDE
• METHYLENE CLORIDE
• ME THYL ALCOHOL.
• ALL KINDS OF ACIDS
Faro has tested and suggests the following disinfectants:
Eco Jet-1 (Cattani Group) / Sporekin Plus DS (Ims srl) / Zefirol Quick (Molteni
Dental) / Durr FD366 Sensitive

ARM CLEANING

Warning against danger of corrosion and mechanical collapse with falling
suspended weights
Never spray chemical substances directly on the arms or joints and their
openings.
Warning against danger of wear of the plastic with falling suspended
weights
D o n ot u se d etergents-disinfectants c ontaining t he f ollowing t o c lean p lastic p arts:
• AMMONIUM HYDROXIDE
• SODIUM HYDROXIDE
• METHYLENE CLORIDE
• ME THYL ALCOHOL.
• ALL KINDS OF ACIDS
Faro has tested and suggests the following disinfectants: Eco Jet-1 (Cattani Group) /
Sporekin Plus DS (Ims srl) / Zefirol Quick (Molteni Dental) / Durr FD366 Sensitive
Warning - danger of cross contamination
The handles are not supplied sterile, they must therefore be sterilised before
use. The handles must be sterilised before each patient.

Removing the handles

Decontamination and disinfectant

Attention - danger of plastic breaking
T he handles cannot be disinfected by thermo-disinfection.
"A"