Connecting
2.2 Mounting the Transmitter
3. Pull either the AC or DC wires through transmitter case cable gland and into the
transmitter case before wiring power connector.
Figure 3-1
Note
Refer to figure above, callouts
Maximum cable lengths cited may be extended with factory approval, if necessary.
4. Locate power plug P12 and loosen P12 connector screws.
12
NOTICE
Do not use a screwdriver to pry up the Display board connector (circled in the figure
below).
Damage to the unit may occur.
①
Fuse F1 Types: GMA, 2A (DC);
500 mA (AC)
②
Input Power Connector P12
③
I/O Terminals TB1
④
J8 RS-485 Port Connector - 30m
(98 ft) max length
Input Power Wiring
⑤
J9 RS-232 Port DB9 Connector - 15m (50 ft) max
length
⑥
S1 Keypad Enable Switch
⑦
I/O Input Cables - 30m (98 ft) max length
⑧
Input Power Cables
⑨
Sensor Cables - 90m (300 ft) max length
④
⑤
⑦
⑨
,
,
and
.
Operating Instructions, 08/2014, A5E03086489-AJ
FST020 IP65 NEMA 4X Quick Start