Connecting
3.3 Installing the Sensor
4. Prepare the two areas you marked by degreasing the surface, if needed, and removing
any grit, corrosion, rust, loose paint or surface irregularities with the abrasive pipe
conditioning material provided. Clean the pipe of all debris and abrasive particles.
Figure 3-8
5. Wrap a Mounting Strap around the pipe. Make sure to position strap so there is easy
access to the Mounting Strap Adjusting Screw.
6. Place the assembled Mounting Frame so that it rests on the top of the pipe.
7. Engage the end of the Mounting Strap with the Mounting Strap Adjusting Screw.
8. Slide Mounting Starp under the spring clip of one of the Mounting Frames.
9. Tighten the Mounting Strap Screw enough to take up all the slack, but not enough to
prevent rotation of the Frame Assembly. Repeat Mounting Strap procedure for the other
Mounting Frame.
10.Rotate and align the Mounting Frame assembly to the final conditioned pipe surface
location, ensuring that the assembly is straight along the pipe axis. (Refer to the sensor
orientation in the Sensor Installation figure below.)
11.Tighten the Mounting Straps to seat the assembly firmly on the pipe. Do not over tighten.
30
①
Clean areas
Preparing the Pipe
FST020 IP65 NEMA 4X Quick Start
Operating Instructions, 08/2014, A5E03086489-AJ