Connecting
3.3 Installing the Sensor
3.3.1
Reflect Mount using Spacer Bar
The spacer bar eliminates manual spacing measurements and provides rigidity for mounting
the sensors while maintaining axial alignment. Refer to Programming the Transmitter
(Page 18) before proceeding.
1. Perform all required menu steps taking note of the sensor model and size designation.
The transmitter then issues the number index and prompts you to press <ENTER> to
finish the sensor install routine. Stop at this point.
Note
Note the number index value displayed in the [Install Sensor] menu. You will use this
index to properly space the sensors. Check to ensure that you have a matched set of
sensors. They both should have the same S/N number but marked with either an "A" or
"B" (e.g., 19256A and 19256B).
2. Prepare the pipe surface area where the sensors will be mounted. Degrease the surface
and remove any grit, corrosion, rust, loose paint, etc.
Note
Before beginning refer to the Reflect Mount sensor installation diagram below.
Figure 3-6
28
①
Mounting Frame
②
Spacer Bar Screw Hole
③
Mounting Strap Adjusting Screw
④
Optional: On larger pipes, multiple
lengths of straps can be linked
together to surround pipe
Reflect Mount with Mounting Frames and Spacer Bar
⑤
Mounting Strap positioned around Mounting Frame
⑥
Locking Nut
⑦
Sensor Clamping Screw
Operating Instructions, 08/2014, A5E03086489-AJ
FST020 IP65 NEMA 4X Quick Start