6 English
Test specifications for enclosure port immunity to RF wireless communications equipment
Test frequency
(MHz)
385
450
710
745
780
810
870
930
1720
1845
1970
2450
5240
5500
5785
a
For some services, only the uplink frequencies are included.
b
The carrier shall be modulated using a 50 percent duty cycle square wave signal.
c
As an alternative to FM modulation, 50 percent pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.
DIR 80025125 Ver D
a
a
Band
Service
MHz
380 - 390
TETRA 400
430 - 470
GMRS 460,
FRS 460
704 - 787
LTE band 13, 17 Pulse modulation
800 - 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
1700 - 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE Band
1, 3, 4, 25;
UMTS
2400 - 2570
Bluetooth,
WLAN, 802.11
b/g/n, RFID
2450,
LTE Band 7
5100 - 5800
WLAN 802.11
a/n
Welch Allyn
Maximum power
b
Modulation
(W)
1.8
b
Pulse modulation
18 Hz
c
2
FM
±5 kHz
deviation
1 kHz sine
b
0.2
217 Hz
b
2
Pulse modulation
18 Hz
b
2
Pulse modulation
217 Hz
b
2
Pulse modulation
217 Hz
0.2
b
Pulse modulation
217 Hz
®
Spot 4400 Vital Signs
Distance (m)
Immunity
test level (V/
m)
0.3
27
0.3
28
0.3
9
0.3
28
0.3
28
0.3
28
0.3
9