Electromagnetic Immunity - Welch Allyn Spot 4400 Vital Signs Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

The Spot 4400 device is intended for use in the electromagnetic environment specified below. The customer or the user
of the Spot 4400 device should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic discharge
±8 kV contact
(ESD)
±15 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power supply
transient/burst
lines
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±0.5 kV, ±1 kV
IEC 61000-4-5
Line- to -line
±0.5 kV, ±1 kV, ±2 kV
Line-to-ground
Voltage dips, short
100 % U
interruptions and
voltage variations on
power supply input
At 0°, 45°, 90°, 135°,
lines
180°, 225°, 270° and
315°
IEC 61000-4-11
100 % U
70 % U
T
Single phase: at 0°
0 % U
; 250/300 cycle 0 %U
T
Power frequency
30 A/m
(50/60 Hz) magnetic
field IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T

Electromagnetic immunity

Compliance level
±8 kV
±15 kV
±2 kV
±1 kV
±1 kV
±2 kV
; 0.5 cycle
0 % U
; 0.5 cycle
T
T
; 1 cycle
0 % U
; 1 cycle
T
T
; 25/30 cycles
70 % U
; 25/30 cycles
T
; 250/300 cycle
T
30 A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the Spot 4400 device requires continued
operation during power mains interruptions, it
is recommended that the Spot 4400 device be
be powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
English 3

Publicidad

Tabla de contenido
loading

Tabla de contenido