Setup and Interconnection
Note
Stryker Endoscopy considers instructional training, or inservice,
an integral part of the 1188HD Medical Video Camera. Your local
Stryker Endoscopy sales representative will perform at least one
inservice at your convenience to help set up your equipment and
instruct you and your staff on its operation and maintenance. To
schedule an inservice, contact your local Stryker Endoscopy
representative after your equipment has arrived.
Setting Up the 1188HD Camera involves three steps:
1.
Setting up the console
2.
Setting up the camera head
3.
Setting up the coupler
Setting Up the Console
Caution
Equipment which employs RF communications may affect the
normal function of the 1188HD Camera. When choosing a
location for the 1188HD Camera, consult the
"Electromagnetic Compatibility" section of this manual to
ensure proper function.
To set up the console, make the following connections:
1.
Connect AC power.
• Connect the AC power cord to the AC inlet on the rear console panel.
• Connect the other end to a hospital-grade outlet.
Warning
Always connect the camera to an appropriate power source,
using a hospital-grade power cord. Loss of AC power will cause
the camera to shut down and the surgical image to be lost.
Caution
Always set up the console in a location that allows adequate
ventilation (airflow) to the console. Insufficient ventilation
may cause the console to overheat and shut down.
Warning
Only connect items to the camera that have been specified for
use with the camera. Connecting incompatible equipment may
cause unexpected results.
To avoid risk of electric shock, this equipment must only be
connected to a supply mains with protective earth.
EN-12