To avoid potential serious injury to the user and patient, and/or damage to this
device, the user must:
1.
Carefully unpack this unit and check if any damage occurred during
shipment. If damage is detected, refer to the "Service and Claims"
section of this manual.
2.
Read this operating manual thoroughly, especially the warnings, and be
familiar with its contents before connecting and using this equipment.
3.
Be a qualified physician, having complete knowledge of the use of this
equipment.
4.
Test this equipment prior to a surgical procedure. This unit was fully
tested at the factory before shipment.
5.
Never use this equipment in the presence of flammable or explosive
gases.
6.
The camera head surface may exceed 41°C (106°F) in operating
conditions with high ambient temperatures and should be handled with
caution.
7.
Avoid dissembling any part of the camera head, as doing so may break
the seals, causing leakage and/or electric shock.
8.
Avoid removing covers on the control unit, as doing so may cause
damage to electronics and/or electric shock.
9.
To avoid risk of electric shock, this equipment must only be connected
to a supply mains with protective earth.
10. Attempt no internal repairs or adjustments not specifically detailed in
this operating manual.
11. Pay close attention to the care and cleaning instructions in this manual.
Any deviation may cause damage.
12. The camera head and coupler are shipped non-sterile. Clean and
sterilize these components prior to first use and after every subsequent
use. Follow the cleaning, disinfection, and sterilization instructions
provided in these instructions.
13. Never sterilize the camera console, because the delicate electronics can
not withstand this procedure.
14. Never autoclave a camera head unless it is marked
Autoclaving regular camera heads will result in permanent device
damage for which Stryker will not be responsible.
15. Disconnect the control unit from the electrical outlet when inspecting
the fuses.
EN-4
.
AUTOCLAVABLE