Problem
Optics are dirty
Blurry picture
Note
If this troubleshooting guide does not resolve the problem, call
Stryker Technical Support at 1-877-478-7953 (inside the U.S.) or
refer to the "Service and Claims" section of this manual.
Possible Solution
• Rotate the scope. If dust particles in the picture rotate,
the dust is located on the scope itself. Follow the
manufacturer's instructions for cleaning the eyepiece
and negative lens.
• If particles in the picture do not move when you rotate
the scope, the particles are located on the coupler or
camera. Remove the scope and clean the window on
the front of the coupler with a dry or alcohol-tipped
cotton swab.
• If dust particles lie between the coupler and camera,
remove the coupler and clean the coupler and camera
windows.
• Ensure all components are completely dry before
reassembling them, or fogging may result.
• Ensure the coupler or C-Mount scope is in focus.
• Increase the enhancement.
• Ensure the specialty switch is not set to FLEXI-SCOPE
unless you are using a flexible scope.
EN-27