Troubleshooting Guide
Troubleshooting Guide
Problems
Wrong pH in System
Positive Control does
Buffer.
not match historic
data.
Acceptor Sink
Buffer(ASB) with
precipitation.
Decomposition of
compounds.
Well is not painted
properly with lipid.
Stirrer disk missing in
the well.
Undetected Pe.
Data Analysis.
10
Possible Cause
Calibrate pH meter.
Thaw ASB to room temperature
Do not reuse ASB for next day
experiment.
Prepare fresh stock solutions.
Ensure that there is enough lipid
for painting in the lipid supply
plate. Adjust pipetting method in
Step 2 if needed.
Inspect donor plate prior to the
experiment to ensure that stirrer
disks are in all wells.
Adjust Detection Threshold in
Refine Permeability Constants
dialog box. Use only if UV
spectrum can be visually
confirmed.
Suggestions
If none of the listed problems
observed, please, refer to
"pION's Guide to Improve
Double-Sink™ PAMPA" or
User's Guide for data results
acceptances.
For data analysis questions,
please, refer to the User's
Guide, Appendix f and "pION's
Guide to Improve
Double-Sink™ PAMPA"
Biomek FX/PAMPA Evolution96 Quick Start and Trouble Shooting Guides