1.6 PRECAUTIONS
• Always test the function of the Provox XtraHME Cassette prior to use. The
top lid should immediately return to its open position after releasing the
finger.
• Do not disassemble the Provox XtraHME Cassette since this will interfere
with its proper function.
• Do not reuse the Provox XtraHME Cassette or attempt to rinse it with water
or any other substance. This will substantially reduce the function of the
HME. Additionally the risk of potential infections may increase due to
bacterial colonization of the foam.
• Do not use the Provox XtraHME Cassette longer than 24 hours. The risk
of potential infections may increase with the time of use due to bacterial
colonization of the foam.
• Do not administer medicated nebulizer treatment over the device since the
medication can become deposited in the device.
• Do not use humidifiers or heated humidified oxygen via a mask over the
tracheostoma while using the device. The HME will become too wet. If
oxygen therapy is required, use only non-heated humidified oxygen.
2. Instructions for use
2.1 Operating instruction
Insert the Provox XtraHME Cassette into the connector of the attachment
device (Fig. 1 or 2). Breathe normally.
To speak, press the top lid of the Provox XtraHME Cassette down with
a finger (Fig. 3).
Note: Always release the lid completely at inhalation to avoid increased
breathing resistance.
To remove the Provox XtraHME Cassette, hold the attachment device in
place with two fingers and remove the HME Cassette from the holder (Fig. 4).
2.2 Device lifetime and disposal
The HME is for single use and must be replaced at least every 24 hours, or more
often if needed.
Always follow medical practice and national requirements regarding
biohazard when disposing of a used medical device.
2.3 Accessories
The Provox XtraHME Cassettes are mainly intended to be used with the other
components of the Provox HME System: the Provox Adhesives, the Provox
LaryTubes and the Provox LaryButtons (See Ordering information).