INSTRUCTIONS FOR USE
e. Once the stirrup is placed in the rail clamp, tighten handle securely.
WARNING: The handle must be tightened securely before use.
f. To pre-position stirrup boots, loosen boot-locking mechanism and slide boot
along rod until top end (calf portion) of boot is located approximately at
patient's mid-calf. Tighten Boot Locking Mechanism securely. Repeat above
procedure for the other stirrup.
g. Position stirrups to be level with the table by squeezing the release handle and
moving stirrup downward.
Positioning Patient in Stirrups
h. Ensure the patient is positioned on O.R. table in accordance with procedure
and surgeon requirements.
i.
Pre-position stirrup boots. Loosen boot-clamping mechanism and slide boot
along rod until top end (calf portion) of boot is approximately at patient's mid-
calf. Tighten clamp.
NOTE: For larger patients, a padded arm shield can be used to protect the wrist,
hand and arm.
Document Number: D-720402
Version: D
Page 15
Issue Date: 18 MAR 2020
Ref Blank Template: 80025117 Ver. F