M-LNC MP Patient Cable
M-LNC MP Series
Reusable
Prior to using this Cable, the user should read and understand the Operator's Manual for the Device, this Directions for Use
and the Directions for Use for the Sensor.
INDICATIONS
The M-LNC MP Series Patient Cables and Masimo sensors are indicated for the continuous, noninvasive monitoring of functional
oxygen saturation of arterial hemoglobin (SpO
neonatal patients in hospitals, hospital-type facilities, mobile, and home environments.
DESCRIPTION
The M-LNC MP Series Patient Cables are for use on Philips modules with Masimo SET® Technology and devices with Philips FAST-SpO
Technology.
On Philips modules with Masimo SET Technology, M-LNC MP Patient Cables are for use with non-rainbow M-LNCS® sensors.
On devices with Philips FAST-SpO
Technology is not available on devices with Philips FAST-SpO
WARNING: Masimo sensors and cables are designed for use with devices containing Masimo SET® oximetry or licensed to use
Masimo sensors.
WARNINGS, CAUTIONS AND NOTES
• Always refer to the oximeter module operator's manual for complete instructions or additional instructions.
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable and discard if
cracks or discoloration are found.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable and sensor before
use, otherwise degraded performance and/or patient injury can result.
• Carefully route cable and patient cable to reduce the possibility of patient entanglement or strangulation.
• Failure to properly connect the sensor or the oximeter module to the cable will result in intermittent readings, inaccurate results,
or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or disconnecting either end.
• To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to sterilize the cable or sensor.
• Do not attempt to reprocess, recondition or recycle Masimo patient cables as these processes may damage the electrical
components, potentially leading to patient harm.
• Refer to the specific sensor's Directions for Use for detailed sensor instructions and warnings.
• Probes and cables are designed for use with specific monitors. The user and/or operator needs to verify the compatibility of the
monitor, probe, and cable before use, otherwise patient injury can result.
• Caution: Replace the cable when a replace cable or equivalent message is displayed consistently. Consult monitoring device
operator's manual for more information.
• Note: The cable is provided with X-Cal® technology to minimize the risk of inaccurate readings and unanticipated loss of patient
monitoring. The cable will provide up to 17,520 hours of patient monitoring time when connected to a monitoring device that
supports X-Cal technology. Replace the cable when the patient monitoring time is exhausted. Some legacy monitoring devices
do not distinguish between the cable and the sensor. In these devices, a replace sensor or similar message could pertain either to
the sensor or the cable.
INSTRUCTIONS
A) Attaching the M-LNC MP Patient Cable directly to the Instrument
Refer to Fig. 1. Orient the connector of the cable to mate with the patient cable connector on the Module.
Caution: Ensure you are connecting a M-LNC MP Patient Cable to either a Masimo SET or Philips FAST-SpO
Module. There are a number of similar connectors with different colors and different mechanical keying. Never force a patient cable
connector into the module. Failure to use a M-LNC MP patient cable may result in damage to the module, inaccurate readings, or
no readings.
Note: There is a difference in color shade between the M-LNC MP Patient Cable and the Philips Module patient cable connector.
However, this is an acceptable configuration. Masimo SET performance is not available when connected to an instrument with
Philips FAST-SpO
Technology.
2
B) Attaching the M-LNC MP Patient Cable to the Sensor
M-LNCS Reusable Sensor
1. Refer to Fig. 2. Orient the reusable sensor connector (1) to the M-LNC MP patient cable (2) as shown.
2. Insert the reusable sensor connector (1) completely into the M-LNC MP patient cable (2). Close the reusable sensor connector
cover (3) over the M-LNC patient cable connector (2).
M-LNCS or rainbow Adhesive Sensor
1. Refer to Fig. 3. Orient the adhesive sensor connector (1) to the M-LNC MP patient cable connector (2) as shown.
2. Insert the adhesive sensor connector (1) completely into the M-LNC MP patient cable connector (2) until it locks in place.

DIRECTIONS FOR USE

LATEX
Not made with natural rubber latex
PCX-2108A
02/13
) and pulse rate (measured by an SpO
2
Technology, M-LNC MP Patient Cables are for use with non-rainbow M-LNCS sensors. Masimo SET
2
2
Technology.
2
3
sensor) for use with adult, pediatric, infant, and
cable connector on the
2
en
Non-sterile
2
5729H-eIFU-0119