Recommended separation distances
Recommended separation distances between portable and mobile RF
communication equipment and the COM-DEX Remote Mic.
The COM-DEX Remote Mic is intended for use in the electromagnetic envi-
ronment in which RF disturbances are controlled. The customer or the
user of the COM-DEX Remote Mic can help prevent electromagnetic inter-
ference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the COM-DEX Remote
Mic as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum
Separation distance according
output power of
to frequency of transmitter
transmitter
m
W
150 kHz to 80
MHz
d = 1.2 √P
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
80 MHz to 800
MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
94
800 MHz to
2.5 GHz
d = 2.3 √P
0.23
0.73
2.3
7.3
23