I-Tech UT2 Manual De Instrucciones página 38

Ocultar thumbs Ver también para UT2:
Tabla de contenido

Publicidad

Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY – FOR
EQUIPMENT AND SYSTEMS THAT ARE NOT LIFE-SUPPORTING
The I-TECH UT2 device is intended for use in. the electromagnetic environment
specified below. The customer or the user of the I-TECH UT2 should assure that it is
used in such an environment.
Immunity test
Portable and mobile RF communications equipment should be used no closer to any
part of the I-TECH UT2 device, including cables, than the recommended separation
distance calculated from the equation applicable to the frequency of the transmitter.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Si può verificare interferenza in prossimità di apparecchi contrassegnati dal seguente simbolo:
Where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey
, should be less than the compliance level in each frequency range
a
Interference may occur in the vicinity of equipment marked with the symbol above.
Note:
(1)
At 80 MHz and 800 MHz, the higher frequency range applies.
(2)
Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
a)
Field strengths from fixed RF transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which I-TECH UT2 is used
exceeds the applicable RF compliance level above, I-TECH UT2 should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating I-TECH UT2.
b)
Over the frequency range 150kHz to 80MHz, field strengths should be less than [V
IACER Srl
Test level
Compliance
IEC 60601
3V
3V
eff
eff
from
150kHz
from
to 80MHz
to 80MHz
10V/m
10V/m
from 80MHz to
[E
] V/m
1
2,7GHz
level
environment – guidance
Recommended separation distance
���������������� 150������������ �������� 80������������
150kHz
���������������� 80������������ �������� 800������������
���������������� 800������������ �������� 2,7������������
38
Electromagnetic
���� = 1.2√����
���� = 1.2√����
���� = 2.3√����
.
b
] V/m.
1
MNPG130-02

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

11248

Tabla de contenido