MB23 / MB25
6. TECHNICAL DATA
Admissible Ambient Conditions
Location:
Temperature:
Relative humidity:
Warm up time:
Height above sea level: Up to 2000 m
Power Input:
Voltage fluctuations:
Power load:
Power line fuse:
Protected against dust and water, Pollution degree: 2, Installation category: Class II
Specifications
MODEL
Capacity
Readability (% requires >10g sample)
Temperature Settings
Heat Source
Calibration - Weight
Calibration - Temperature
Pan Size
Unit Dimensions (DxWxH)
Unit Weight
Shipping Dimensions (LxWxH)
Shipping Weight
Communication
RS232 Pin Connections
GND
5
9
RS232 Data Settings
Baud Rate (default ): 2400 Data Bits: 7
Baud Rate:
RS232 Commands
The RS232 Interface allows a computer to control the Analyzer, as well as to receive data such as displayed weight.
Command
ON
OFF
T
U
START
STOP
P
xP
PSN
PV
?
H
RS
Indoor use only
10° C to 40° C
15 % to 80 % at 30° C non-condensing
At least 15 minutes after connecting the Analyzer to the AC supply;
when switched from standby mode the Analyzer is ready for immediate use.
100VAC - 120VAC, 3A, 50/60Hz or 200VAC -240VAC, 3A, 50/60Hz
-15% +10%
MB25: 250W; MB23:300W (Maximum during drying process)
1 piece, 5 x 20mm, 2.5A 250V
Female DB9 connector
RxD
TxD
Pin 2: Analyzer transmit line (TxD)
1
Pin 3: Analyzer receive line (RxD)
6
Pin 4 and 6: Internally connected
Pin 5: Ground signal (GND)
CTS
Pin 7: Clear to send (hardware handshake) (CTS)
RTS
Pin 8: Request to send (hardware handshake) (RTS)
Parity: N
9600 Data Bits: 8
Parity: N
Function
Turns Analyzer ON
Turns Analyzer OFF
Same as pressing Tare
Same as pressing %g
begins a test
ends a test
Same as pressing Print
Interval Print x = Print Interval (1-3600 sec)
Print Serial Number
Print software version
Print Header
Header on or off
Print current RS232 settings
MB25
110g
0.005g ; 0.05%
50° C to 160° C (5° increments)
Halogen
50g
OHAUS Temperature Calibration Kit
90mm diameter
11.0x6.5x5.0 in / 28.0x16.5x12.7 cm
4.6 lb / 2.1 kg
20.2x15.5x13.5 in / 51x40x35 cm
10 lb / 4.5 kg
Stop Bits: 2
Stop Bits: 1
MB23
0.01g, 0.1%
Infrared
Flow Control: Xon/Xoff
Flow Control: Xon/Xoff
EN-9