• The lip clip and the hook must be sterilized between treatments. Please note that the
measuring cable cannot be autoclaved.
• In addition, a fork is not included in the propex
follow the same procedure as the lip clip and hook.
• Follow the "Disinfection and sterilization procedure" described in section 7.2.
• The user is responsible for the sterility of the lip clip, the connection hook and the
fork for the first cycle and each further usage.
• All damaged accessories should be discarded and dirty accessories should be clea-
ned and sterilized per the procedure described in section 7.2
7.2 Disinfection and sterilization procedure for lip clip,
connection hook and fork
Foreword
For hygiene and sanitary safety purposes, the lip clip, the connection hook and the fork
must be cleaned, disinfected and sterilized before each usage to prevent any cross-
contamination between patients. This concerns the first use as well as the subsequent
uses.
General recommendations
1 - Use only a disinfecting solution which is approved for its efficacy (VAH/DGHM-lis-
ting, CE marking, FDA approval) and in accordance with the DFU of the disinfecting
solution manufacturer. For all metal instruments, it is recommended to use anticorro-
sion disinfecting and cleaning agents
2 - For your own safety, please wear personal protective equipment (gloves, glasses,
mask).
3 - The user is responsible for the sterility of the product for the first cycle and each fur-
ther usage as well as for the usage of damaged or dirty instruments where applicable
after sterility.
4 - Limitations and restrictions on reprocessing:
the appearance of defects such as cracks, deformations (bent, twisted), corrosion, loss
of colour coding or marking, are indications that the devices are not able to fulfil the
intended use with the required safety level.
5- Use only clean water in all cleaning and rinsing steps.
pixi
but it can be used and should