For transmitters operating below 150 kHz (RFID devices such as access control devices), the recommended separation distance is at least 1.17 m.
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
The Precision Montage MRI System is intended for use in the electromagnetic environment specified below. The customer or user of the Precision
Montage MRI System should assure that it is used in such an environment.
Emissions Test
Compliance
RF emissions CISPR 11
Group 1
RF emissions CISPR 11
Class B
Harmonic emissions
Class B
IEC 61000-3-2
Voltage fluctuations / Flicker
Complies
emissions IEC 61000-3-3
Hereby, Boston Scientific declares that the Precision Montage MRI System is in compliance with theessential requirements and other
relevant provisions of Directive 1999/5/EC.
Precision montage™ mri system information for Patients
91053248-04 reV a
22 of 679
Guidance and Manufacturer's Declaration - electromagnetic emissions
Electromagnetic Environment Guide
The Precision Montage MRI System uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
The Precision Montage MRI System is suitable for use in all establishments. Including
domestic establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.