4 Technical Data
Table 2 Measurement Performance
Property
Hydrogen
Measurement range (in oil)
Accuracy (in oil temperature range
-20 ... +60 °C (-4 ... +140 °F))
Moisture in Oil
Measurement range (in oil)
Response time (90 % of full response at +20 °C
(+68 °F) in still oil)
Sensor
Accuracy (including non-linearity, hysteresis, and repeatability):
0 ... 90 %RS
90 ... 100 %RS
Temperature
Measurement range
Accuracy at +20 °C (+68 °F)
1)
Note that the temperature range for specified measurement accuracy differs between
measurement parameters. Measuring a parameter outside its temperature range does not
damage the sensor, but results in reduced accuracy for that parameter.
Table 3 Operating Environment
Property
Oil type
Operating temperature (electronics)
Storage temperature
Operating humidity
Description/Value
0 ... 5000 ppm
±20 % of reading or ±25 ppm
1)
greater)
0 ... 100 %RS / a
10 min
HUMICAPâ 180L2
±2 %RS (a
±3 %RS (a
-40 ... +120 °C (-40 ... +248 °F)
±0.2 °C (0.36 °F)
Description/Value
Selected when ordering, available options:
• Mineral oil
• Natural ester oil
• Synthetic ester oil
-40 ... +60 °C (-40 ... +140 °F)
-40 ... +60 °C (-40 ... +140 °F)
0 ... 100 %RH, condensing
Chapter – Technical Data
v
(whichever is
v
0 ... 1
w
± 0.02)
w
± 0.03)
w
1)
1)