Technical information

Type: TA01i.
The device conforms to CE regulations.
Dimensions: 13 – 32 mm (W) * 45 – 80 mm (H) * 230 mm (L).
Weight: 155 g (without batteries), 250 g (4 x AA batteries).
Power supply: 4 x AA batteries.
Measurement range: 7-50 mmHg, display range: 0-99 mmHg (IOP estimation beyond
the measuring range).
Accuracy (95% tolerance interval relative to manometry): ±1.2 mmHg (
and ±2.2 mmHg (>20 mmHg).
Repeatability (coefficient of variation): <8%.
Accuracy of display: 1.
Display unit: Millimeter mercury (mmHg).
The serial number is on the back of the battery compartment cover.
There are no electrical connections from the tonometer to the patient.
The device has B-type electric shock protection.
Storage/transportation environment:
Temperature +5 to +40 °C.
Rel. humidity 10 to 80 % (without condensation).

Spare parts and supplies

Single-use probes.
Probe base replacement kit.
WARNING -Indicates that important operating instructions are included in this
Instruction Manual.