Technical information
•
Type: TA01i.
•
The device conforms to CE regulations.
•
Dimensions: 13 – 32 mm (W) * 45 – 80 mm (H) * 230 mm (L).
•
Weight: 155 g (without batteries), 250 g (4 x AA batteries).
•
Power supply: 4 x AA batteries.
•
Measurement range: 7-50 mmHg, display range: 0-99 mmHg (IOP estimation beyond
the measuring range).
•
Accuracy (95% tolerance interval relative to manometry): ±1.2 mmHg (
and ±2.2 mmHg (>20 mmHg).
•
Repeatability (coefficient of variation): <8%.
•
Accuracy of display: 1.
•
Display unit: Millimeter mercury (mmHg).
•
The serial number is on the back of the battery compartment cover.
•
There are no electrical connections from the tonometer to the patient.
•
The device has B-type electric shock protection.
•
Storage/transportation environment:
•
Temperature +5 to +40 °C.
•
Rel. humidity 10 to 80 % (without condensation).
Spare parts and supplies
•
Single-use probes.
•
Probe base replacement kit.
WARNING -Indicates that important operating instructions are included in this
Instruction Manual.
≤20 mmHg)