Aesculap
Pneumatic motor GA553
8.
Troubleshooting list
Malfunction
Motor is not running
Insufficient power
9.
Technical Service
Risk of injury and/or malfunction!
►
Do not modify the product.
WARNING
►
For service and repairs, please contact your national B. Braun/
Aesculapagency.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
10
®
Detection
Gas cylinder empty
Pressure Gauge set to "0"
No compressed air at the wall outlet
The ring for rotation direction selec-
tion 22 is on "0"
Hoses are not correctly connected
Pedal can be easily pressed or cannot
be pressed at all
The ring for rotation direction selec-
tion 22 is not on the end stop
Operating pressure of the central com-
pressed air supply too low (min. 6 bar)
Pressure reducer on gas cylinder defec-
tive
Pressure on pressure reducer set too
low
Hose is too long
Hoses excessively kinked or crushed
Exit air does not flow freely
Motor does not run in spite of correct
compressed-air supply
Pneumatic motor heats up
Pneumatic motor runs jerkily
Cause
No compressed air
Foot control defective
Compressed air supply pressure
too low
High air resistance in the hose
Pneumatic motor faulty
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Remedy
Fill or change cylinder
Open valve
Turn the ring for rotation direction
selection 22 to the right or the left
Fully connect the hose connector
Have product repaired by the manufac-
turer
Turn the ring for rotation direction
selection 22 until it clicks on the stop
(preferred direction of rotation: clock-
wise)
Have compressed-air supply checked
Have pressure reducer checked
Set higher pressure
Use shorter hoses
Do not kink or squeeze hoses
Diffuser 14 or exhaust air channel
Have product repaired by the manufac-
turer
Have product repaired by the manufac-
turer
Have product repaired by the manufac-
turer