TROUBLESHOOTING
Symptom
Pressure too low
Suction too low at the Swiss LithoPump
The Swiss LithoPump jolts in operation
Suction level cannot be reached
Suction too low with a ultrasound probe
Suction too low at the pneumatic probe
The Swiss LithoPump is noisy
46
FB-350_3_ed_2010-06_Lithopump.indd 46
Solution
•
Check system air supply.
•
Check the compressor if applicable.
•
Check that air arrives at the suction outlet (large connector).
•
Check that air suppply pressure is more than 5 bar.
•
Check the mounting orientation of the internal filter.
•
Check that the filter is not obstructed or moist, replace the filter.
•
Adjust the suction level.
•
The air exhaust is obstructed.
•
Stop the Swiss LithoPump to allow pressure to reach its maximum output (if applicable).
•
Possible overheat of the compressor when a compressor is used.
•
Use a ultrasound probe with a larger diameter to improve suction.
•
Check the suction at the hose connected to the handpiece.
•
Check the suction at the hose connected to the handpiece.
•
If the suction is OK, verify the blue colored seals between the handpiece and the LithoVac.
•
Verify the circular seal on the LithoVac probe.
•
Verify of the suction line up to the Swiss LithoPump.
•
Replace the silencer.