13. Troubleshooting Guide

Problem
No air and no abrasive.
Only air, no or very
little abrasive.
Unit blasts continu-
ously.
Compressed air
leaking from the foot
switch.
Foot switch fails to
switch on.
EN
Possible cause
Blasting nozzle blocked at the
hand-piece.
Foot switch connecting lines
kinked.
The dosage nozzle in the mix-
ing chamber is blocked due to
contaminated or moist sand.
• Mixing chamber filter combination
blocked.
Abrasive contains impurities or is
moist.
Foot switch connected incorrectly.
Valve in foot switch is defective.
Pneumatic hoses reversed.
No pressure at the compressed
air connection.
Pneumatic hoses are kinked.
Corrective action
Unscrew the nozzle from the hand-piece and
blow it out.
With the nozzle removed, blow out the blasting
hose by activating the foot pedal.
Inspect/correct the connection lines.
Cleaning the dosing nozzle, see chapter 5.4.
• Change filter cartridge (Wrench SW14).
Replace the abrasive.
Inspect / correct the foot switch connection.
Replace foot switch.
Reverse the connections.
Inspect the compressed air lines and/or the com-
pressor.
Reposition the hoses.