Packaging; Sterilization - Erbe BiClamp 110/340 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 52
OBJ_DOKU-163748-015.fm Seite 32 Donnerstag, 10. Dezember 2020 1:48 13
7.9

Packaging

The product can be sterilized while it is dismantled or assembled.
The jaws must be open.
The product must be protected against damage during steriliza-
tion.
1. If required, assemble the product. See page 24.
2. Pack the product in disposable sterilization packaging (single
or double packaging) made from paper/foil and/or in a ster-
ilization container.

7.10 Sterilization

Only sterilize products that have been cleaned and disinfected.
Erbe Elektromedizin recommends steam sterilization with the
method described below. If other sterilization methods are used,
Erbe Elektromedizin shall bear no responsibility.
Steam sterilization
• Fractionated vacuum method with adequate product drying
• Holding time 3 to 18 minutes at 132 to 138 °C
• Sterilizers in accordance with applicable national standards
and regulations (e.g. DIN EN 13060 or DIN EN 285)
• Sterilization process validated according to DIN EN
ISO 17665
Please follow the recommendations issued by the sterilizer man-
ufacturer concerning loading, handling, and drying times.
32

Publicidad

Tabla de contenido
loading

Tabla de contenido