Recommended separation distances
P
is intended for use in an electromagnetic environment in which radiated radiofrequency
IEZOTOME
disturbances are controlled.
The P
user and/or installer can help prevent electromagnetic interference by maintaining a
IEZOTOME
minimum distance between portable and mobile radiofrequency communications equipment
(transmitters) and P
recommended in the table below.
Rated max. power of the
transmitter (W)
0.01
0.1
1
10
100
For transmitters rated at max. power not listed above, the recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the max. power rating of the emitter in watts (W) according to the
manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and persons.
P
cable lengths
IEZOTOME
Cables and accessories
Handpiece cord
Control footswitch
, according to the maximum output power of the equipment, as
IEZOTOME
Separation distance in metres (m) according to transmitter frequency
150 KHz to 80 Mhz
P
d = 1.2
0.12 m
0.38 m
1.2 m
3.8 m
12 m
Maximum length
2 m
80 MHz to 800 MHz
P
d = 1.2
0.12 m
0.38 m
1.2 m
3.8 m
12 m
RF emission, CISPR 1 - Class B/Group 1
Harmonic current emission - IEC 61000-3-2
Voltage fluctuation - IEC 61000-3-3
Immunity to electrostatic discharge - IEC 61000-4-2
Immunity to electrical fast transients/bursts - IEC 61000-4-4
Immunity to surges - IEC 61000-4-5
Immunity to Voltage dips, short interruptions and voltage variations -
IEC 1000-4-11
Immunity to conducted disturbances induced by radiofrequency fields -
IEC 61000-4-6
Immunity to radiated radiofrequency electromagnetic fields - IEC 61000-4-3
11
800 MHz to 2.5 GHz
d = 2.3
0.23 m
0.73 m
2.3 m
7.3 m
23 m
Complies with
P