Satelec PIEZOTOME Manual De Uso página 21

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 44
FAULT
The control screen
is lighting up but
neither the
Footswitch disconnected or faulty
ultrasound nor the
irrigation systems
are working
Tip not screwed in properly
Tip deformed or worn
Ultrasound too
Handpiece and/or tip not being used
strong or too weak
properly
Fluid present in the N
handpiece connector
Tip not screwed in properly
Handpiece faulty
Oxidation and/or a black residue
around the N
No ultrasound
output
Cord cut
Handpiece not suitable
Fluid present in the N
handpiece connector
Fluid leakage in
the N
Small water-proof washer in the
EWTRON
handpiece
N
EWTRON
connector
Fluid leakage at
Disposable pump cartridge damaged Change the pump cartridge (see Section 4.3.c)
the end of the tip
TR OR TO
DISPLAYED and no
Internal problem with the device's
ultrasound output
functioning
or irrigation
POSSIBLE CAUSE
EWTRON
connector pins
EWTRON
EWTRON
handpiece worn out
Check that the footswitch is properly connected
Contact the After-Sales Service of SATELEC see Section
10.2)
Screw the tip back in. Refer to the manual provided
with the equipment
Replace the tip. Refer to the manual provided with
the equipment
Refer to the manual provided with the equipment
Dry out the connector. If there is a black residue
around the connector pins and/or the handpiece
terminals, contact the SATELEC After-Sales Service
(see Section 10.2)
Screw the tip back in. Refer to the manual provided
with the equipment
Replace the handpiece. Refer to the manual provided
with the equipment
Clean the contacts, replace the cord or contact the
SATELEC After-Sales Service (see Section 10.2)
Replace the cord
Refer to the manual et provided with the equipment
Dry out the connector. If there is a black residue
around the connector pins and/or the handpiece
terminals, contact the SATELEC After-Sales Service
(see Section 10.2)
Replace the washer. Refer to the manual provided with
the equipment
Turn the device off and then turn it on again. If the
problem persists, contact the SATELEC After-Sales
Service (see Section 10.2)
21
REMEDY

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido