Satelec PIEZOTOME Manual De Uso página 9

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 44
Certain types of mobile telecommunication devices such as mobile telephones are likely to interfere
with P
.
IEZOTOME
The separation distances recommended in the "Electromagnetic Compatibility" chapter must be
complied with.
P
should not be used adjacent to or stacked with other equipment.
IEZOTOME
If adjacent or stacked use is necessary, the Piezotome should be observed to verify normal operation
in the configuration in which it will be used.
The use of accessories other than those specified or sold by Satelec as replacement parts may increase
emissions or reduce the immunity of P
Electromagnetic emissions
P
is intended for use in the electromagnetic environment specified in the table below.
IEZOTOME
The user and/or the installer must ensure that P
Emission test
RF emission CISPR 11.
RF emission CISPR 11.
Harmonic current emission
IEC 61000-3-2.
Voltage fluctuation and flicker
IEC 61000-3-3.
Electromagnetic immunity
P
is intended for use in the electromagnetic environment specified in the table below.
IEZOTOME
The user and/or installer must ensure that P
Immunity test
Electrostatic
± 6 KV contact
discharge (ESD)
± 8 KV air
IEC 61000-4-2.
Electrical fast
transients
± 2 KV for power supply lines
IEC 61000-4-4.
Surges
± 1 KV differential mode
IEC 61000-4-5.
± 2 KV common mode
<5% U
(>95% dip in U
T
Voltage dips, short
cycles
interruptions and
40% U
(60% dip in U
T
70% U
(30% dip in U
voltage variations
T
<5% U
(>95% dip in U
IEC 61000-4-11.
T
cycles
IEZOTOME
Compliance
P
IEZOTOME
Group 1
emissions are very low and are not likely to cause any interference in nearby equipment.
Class B
P
IEZOTOME
Class A
those directly connected to the public low voltage power supply network that supplies
buildings used for domestic purposes.
Complies
IEC 60601 test level
) for 0.5
T
) for 5 cycles
T
) for 25 cycles
T
) for 250
T
.
is used in such an environment.
IEZOTOME
Electromagnetic environment - guidance
uses RF energy only for its internal function. Therefore, its radiofrequency
is suitable for use in all establishments, including domestic establishments and
is used in such an electromagnetic environment.
IEZOTOME
Compliance level
± 6 KV contact
± 8 KV air
± 2 KV for power supply lines
± 1 KV differential mode
± 2 KV common mode
<5% U
(>95% dip in U
) for 0.5
T
T
cycles
40% U
(60% dip in U
for 5 cycles
T
T
70% U
(30% dip in U
) for 25 cycles
T
T
<5% U
(>95% dip in U
) for 250
T
T
cycles
9
Electromagnetic environment - guidance
Floors must be wood, concrete, cement or
tiled. If the floors are covered with synthetic
material, the relative humidity must be at least
30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the use
of Piezotome requires continued operation
during power mains interruptions, it is
recommended that the product be powered
from a separate power supply (UPS, etc.).

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido