2.2.10   Discharge valve installation 
Note: Please open the valve (14) before install it in case of damage the vessel body. 
1)RE‐610 discharge valve (14) installation 
Put  80#  flange  mouth  screw  (59)  and  80#  flange  mouth  retainer  ring  (60)  on  the  flange  mouth  of  vessel  body  (9)  and 
discharge  valve  (14),  put  80#  flange  mouth  sealing  ring  (61)  on  the  flange  mouth  of  discharge  valve  (14),  position  it  in 
place, tighten the butterfly nut.
2.2.11  
Vacuum gauge and control box installation 
Please see chapter 1.4 to find the installation position of Vacuum gauge (4) and control box (17).
2.2.12  
Vessel jacket circulating pipe kit installation 
The lower connecting joint of the vessel jacket is inlet, the upper is outlet. Please refer to figure 2.2.10.2 to install it. 
 
                                           
46
Figure 2.2.10.1 RE‐610 discharge valve installation 
 
 
 
 
 
Figure   
2.2.12   
9      Vessel 
14    Discharge valve 
59    80# flange 
60    80# flange mouth retainer ring 
61    80# flange mouth sealing ring 
 
       
62    Joint 
63    DN15 high temperature ball valve 
64    Square three‐cavities 
65    U shape fixing frame 
66    Gasket 
67    Stainless steel hose 
68    Elbow 
Circulating pipe kit installation 