o) Shipping link
Shipping link and logic interconnections
All logic control inputs and outputs are accessible at
logic connector J22. Ensure that the mains circuit
breaker is off, then remove the provided mating
connector P22 and make the connections with 26 AWG
(or smaller) wire to the pins as indicated below to
obtain the desired capability.
NOTE
With the remote STARTA..OW SPEED connection, the
front panel switches are not operative and viceversa
Input to the Turbo-V controller
PIN
DESCRIPTION
1-9
Remote ST ART optically isolated from
the
internal
permanent logic level of 5 Vdc ( 1 mA
minimum ) and it is TTL compatible.
When this signal is present the Turbo
pump starts and when the signal is
removed the Turbo pump is stopped.
14-5
Remote START requires a permanent
closed contact. If the contact is closed
the Turbo pump starts and stops when
the contact is open.
LOW
7-6
Remote
isolated
from
requires a permanent logic level of
5Vdc (1 mA minimum ) and it is TTL
compatible.
When this signal is present the turbo
pump runs at low speed and when is
not present the Turbo pump runs at
high speed
12-4
LOW
Remote
permanent close contact.
If the contact is closed the Turbo
pump runs at low speed and if the
contact is open it runs at high speed.
13-5
INTERLOCK Ce.a. vacuum pressure
8
+ DC pump current
-r:1---1--- {0. 1 ..-=< 4 A)
o--�.+-.z.. + 24V
60 mA
�'---!--- Relay contact
,-o.::.i--- (IA 250 Y)
doses wilch
pump overload
b) Logic interconnection
circuit
requires
a
SPEED
optically
the
internal
circuit
SPEED requires a
TECHNICAL INFORMATION
switch, water flow switch, feedback
from the forepump relay) requires a
permanent closed contact.
This signal can be used to delay the
starting of the Turbo pump and when
the contact is ooen it stoos.
NOTE
Pin 13-5 must be shorted to allow the Turbo-V pump to
start if a vacuum pressure switch or an external contact
is not connected.
Output from the Turbo-V controller
DESCRIPTION
PIN
11-3
Remote NORMAL speed indication
output (not optically isolated output pin
11 +24 Vdc, pin 3 negative). When the
pump is not in normal speed (either
High or Low speed) the controller
output is an open collector transistor.
15-8
Analog output of
draw 0.1 V proportional to 4 A
2-10
Fault contact.
This contact (which carries 1 A, 250 V
max.) is open when the Turbo pump is
running and closes when the pump is
turned off for a pump / controller
malfunctions condition only.

CONTROLLER TO PUMP CONNECTION

Before connecting the Turbo-V pump to the controller,
ensure correct grounding between pump and controller.
A five-meter long cable is to connect the controller to
the pump. The following figures show the controller
output connector configuration where pins:
A-F = pump temperature sensor
B-G-E = 90 Vac 3-phase output to pump motor
stator
C-D = Spare provision
Controller to pump connector
DC pump current
WARNING!