TECHNICAL INFORMATION
CONTROLLER OUTLINE
The outline dimensions for the Turbo-V 1800 controllers are shown in the following figures.
INTERCONNECTIONS
WARNING!
High voltage developed in the controller can cause
severe injury or death. Before servicing the unit, turn
off rear panel circuit breaker and disconnect input
power cable.
NOTE
The Turbo-V controller can be
has a rack module, but must be positioned so that free
air can flow through the holes.
Power Interconnections
P2l
vent device --------1---,
(120 Vac) __,.'---t-!+I
Power socket output signals
2112.a f\UJ
� -- ---+Ji
,�
j � o
□
ID
Controller models 969-9461 and 969-9561 outline
used
has a bench unit or
Heater Band
N
, ___ l _ (mains voltage)
L
CJ
D
!
DD
The controller is provided with a power connector plug
P21.
Ensure that the main circuit breaker is off, then remove
the plug and wire the pins (maximum wire size is 18
AWG) has indicated below to obtain the desired
capability
PIN
1a
Ground
1b-3a
120 Vac output ( disregarding the
3b-3a
mains} 0.4 A total for vent device and
2b-2a
forepump relay coil.
Voltage is present after ST ART push
button is pressed and remains present
until and overload or until power is
removed from the Turbo oumo.
4b-4a
Output mains voltage 3 A max. for
heater band. Voltage is present when
the Turbo pump is running and the
HEATER oush button is pressed.
jj
DESCRIPTION