Table 4 – Recommended Separation Distances Between Portable And Mobile
RF Communications Equipment And The ME EQUIPMENT Or ME SYSTEM
For ME EQUIPMENT And ME SYSTEMS That Are Not LIFE-SUPPORTING
Recommended Separations Distances for the EGA-500/EGA-500E
The EGA-500/EGA-500E is intended for use in the electromagnetic environment in which radiated
disturbances are controlled. The customer or user of the EGA-500/EGA-500E can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment and the EGA-500/EGA-500E as recommended below, according to the
Max Output
150kHz to 80MHz
Power
(Watts)
D=(3.5/V1)(Sqrt P)
0.01
0.1
1
10
100
Type
Power cable
IFU-51 / REV G
maximum output power of the communications equipment.
Separation (m)
.1166
.3689
1.1666
3.6893
11.6666
Approved accessories
Manufacturer/Model
Interpower/86610610
Separation (m)
80 to 800MHz
D=(3.5/E1)(Sqrt P)
.1166
.3689
1.1666
3.6893
11.6666
EN - 18
Separation (m)
800MHz to 2.5GHz
D=(7/E1)(Sqrt P)
.2333
.7378
2.3333
7.3786
23.333
Maximum length