VI. Front Panel Controls/Connections
NOTE:
Note: All status indicator lights will be green during normal operation.
1. On/Off Push Button
Turns main power on or off to the unit
2. Water Bottle Warmer On/Off
Push to turn water warmer on or off
3. Pump Flow Control
The flow control knob is a continuous variable flow control that has settings between 0 and 100% flow.
The lowest knob setting corresponds to no flow while the max flow control setting represents 100% flow. It
contains bracketed visual iterations to represent flow magnitude in increasing increments of approximately 10%,
however the flow control knob response is not entirely linear. Incremental flow control knob increases between
50-100% have a larger impact on flow rate response as compared to incremental changes between 0-50%.
Flow Rates:
Auxiliary Water Channel
Biopsy Channel
*These values are approximate and are based on an average flow obtained with three different
manufacturer's GI endoscopes with biopsy channel diameters ranging from 2.8 mm-4.2 mm and auxiliary
water channel diameters ranging from 0.8 mm-1.0mm. User results may vary depending on scope used,
channel size and working channel length.
CAUTION:
The ENDO STRATUS® irrigation pump is intended for use only by physicians/clinicians
who are trained regarding the amount of water to use for irrigation and infusion during
GI endoscopic procedures.
4. Foot Pedal Connection
5. Prime Button
When pressed, pump will run for a period of 20 seconds. The cycle can be canceled at any time by either pressing
the foot pedal or prime button again. See Section VIII for full explanation.
6. Pump Head
Refer to Section VIII for assembly instructions
2
1
3
4
(The appearance of your device may differ slightly from picture shown above)
0-300 ml/min*
0-650 ml/min*
5
EN - 9
IFU-51 / REV G